Resource Details
TGP3100 Series Firmware Update V1.08
This zip file contains the firmware (.hex) files for version 1.05 of the firmware for TGP3151, TGP3152, TGP3121 and TGP3122 Pulse and Universal Function Generators. It also contains a PDF readme file with details on how to perform the upgrade.
Version 1.02 included LXI/LAN code updates to fix mDNS network dropout issues.
Version 1.03 added:
- Bug fix "power-on latest" settings for two channel versions
- New Feature: pulse widths of two pulses in double pulse mode can now be set independently
Version 1.04 updates the USB stack to make improvements in the USB CDC communication.
Version 1.05 fixes a firmware problem which prevented both channels of dual channel units being triggered simultaneously via remote control
Versions 1.06 to 1.08 are for internal manufacturing, test and calibration improvements and provide no new features
The current firmware version installed in the generator is shown when the instrument is switched on.
Link: TGP31xx_Firmware_Update_V1.08.zip