We would like to use Google Analytics to be able to monitor how our site performs and how it is used.This requires a few cookies to be set on your device. If you're OK with that, please just continue as normal.If you'd prefer that we didn't, please just turn cookies off using the button below. The Privacy Policy has more information. We'll also be setting a few essential functional cookies to enable the cart, language and currency settings - these are not affected by this setting.

By using our site you accept the terms of our Privacy Policy.

(One cookie will be set to store your preference)
(Ticking this sets a cookie to hide this popup if you then hit close. This will not store any personal information)

About this tool

About Cookie Control
Measurably better value

🛒 Empty

tooltip icon

Resource Details

File/link type icon for ZIP

TGF3XX2 Series Firmware Update V1.03



IMPORTANT: Updating from version 1.00 can effect the instrument calibration. Please read carefully before carrying out the firmware update.

Version 1.01 added two more calibration steps to calibrate the DAC offset in both channels.  These two steps were added to improve the accuracy of the DC offset. 
Version 1.02 fixes a problem which prevented both channels being triggered simultaneously via remote control
Version 1.03 fixes a problem with binary transfers over USB

If updating from V1.00, following the update, the instrument will need to be recalibrated. If recalibration is not performed after updating the firmware, the instrument would perform as before using the current calibration values, however, there will be a message that will pop up every time the instrument is powered on to inform the user that the instrument is not calibrated.

It is recommended that the instrument be recalibrated after updating the firmware from V1.00 to 1.01 or later. The DC offset calibration steps can be performed from the front panel of the instrument with a DC voltmeter without disturbing the existing calibration points.  Instructions are provided in the operation manual.  Alternatively it can be sent off to a calibration facility if traceability is required.   If you do not need improved DC offset accuracy or cannot re-calibrate the instrument, it is recommended that you do not apply this update until calibration is required.

The download is a zip archive containing the firmware files and instructions for updating the instrument.

Language(s): English
Resource Type: Firmware File
Link: TGF3xx2_Firmware_Update_V1.03.zip